program names, file names
variable names
prompts, commands, program code
The filter wheel in the University of Rochester 3rd Generation Infrared Camera has six broadband and three narrowband filters. Each of the narrowband filters is a circular variable filter (CVF), such that each point along a CVF acts as a narrowband filter (~2% resolution at near-infrared wavelengths), and transmission wavelength is a function of position along the filter. Click here to see a schematic diagram of the filter wheel. Below is a table giving approximate characteristics of the three CFVs:
name | minimum wavelength* | maximum wavelength* | short-wavelength end (filter wheel position) |
---|---|---|---|
CV0 ("J-H band") | 1160 nm | 2270 nm | 7223.88 |
CV1 ("K band") | 1850 nm | 2790 nm | 4446.08 |
CV2 ("L band") | 2325 nm | 4575 nm | 1668.28 |
*approximate value |
The filter wheel is rotated by a small electric motor (the "stepper motor") mounted on the side of the camera. The stepper motor moves the filter wheel by angular steps called "filter wheel positions". The last column of the above table gives the filter wheel position of a particular point on each CVF. 10,000 filter wheel positions corresponds to a full 360° rotation. Each CVF has an angular length of 90°, so it takes 2500 filter wheel positions to move from one end of a CVF to the other end.
For the stepper motor and filter wheel to work properly, the computer which operates the camera must know the orientation of the filter wheel at a given moment, and by how much to rotate the filter wheel in order to put the correct filter into the camera's optical path. The filter wheel must therefore be carefully calibrated before observing. We generally do a thorough calibration in the lab before leaving for the observatory, with occasional spot-checks of the alignment during the observing run.
The initial alignment is done using a hydrogen lamp -- a long glass tube filled with hydrogen gas which is stimulated electrically. The lamp is placed under the camera such that the tube is as near as possible to the focus of the camera lens -- this takes a bit of fiddling, but here are a couple of rough estimates: there should be about 10" between the hydrogen lamp tube itself and the dewar baseplate when using the WIRO lens, and there should be about 2.5" between the hydrogen lamp tube itself and the dewar baseplate when using the MLOF lens.
The idea here is to take lots of images of the hydrogen lamp at a sequence of wavelengths such that an atomic transition of hydrogen falls within this wavelength interval. Then, knowing the rest wavelength of the transition, and making a plot of intensity v. filter wheel position, you can determine the filter wheel position corresponding to the wavelength of the hydrogen transition (the intensity increases as you go through the transition, and then drops back down after passing it).
We usually start about 100 filter wheel positions away from the nominal filter wheel position of the transition, and then walk the filter wheel through the transition (we typically step the filter wheel by 10 positions between images). This should be done as quickly as possible, because the glass of the tube gets very hot, and this complicates doing the photometry. It's a good idea to walk the filter wheel through one transition, and then to turn off the lamp and wait for a few minutes before starting again. And, for each set of images, it's important to go far enough past the atomic transtition that the intensity profile falls back down near the baseline value.
Images of the lamp should look something like this:
We usually do photometry at three points along the tube, as seen by the lower three boxes on the gas-filled part of the tube (the "source boxes"). Additionally, we need to do a background subtraction with photometry boxes (the "background boxes") near the source boxes, such that the background boxes are away from the hydrogen gas but still on the glass tube. There are utilites in the data-acquisition software that do this photometry for you. These utilities return box totals, rather than box means, so make sure that each background box is the same size as its corresponding source box (otherwise you'll have to rescale the box totals according to box size).
Once you've got all the box totals, do the background subtractions and make
a plot of intensity v. filter wheel position for each. Then you can just
use
fwhm to find the peak's centroid. You should
get something like this:
The software used to rotate the filter wheel assumes a linear relationship
between wavelength and filter wheel position. So, ideally, you should do
two wavelengths for each CVF (because two points determine a line). This
is possible for CV0 and CV1, but the
backgroud emission gets so high at longer wavelengths, that (in my
experience) there is only one sufficiently bright hydrogen transition that
falls within the wavelength range of CV2 and can be
measured with the hydrogen lamp we use here in the lab. Below is a table
of the transitions we typically use:
transition | CVF | approximate filter wheel position |
wavelength |
---|---|---|---|
Paschen beta | 0 | 6960 | 1282 nm |
Brackett gamma | 0 | 4960 | 2166 nm |
Brackett gamma | 1 | 3610 | 2166 nm |
Brackett beta | 1 | 2390 | 2626 nm |
Brackett beta | 2 | 1330 | 2626 nm |
Let's assume that after taking the data (and after taking the average
filter wheel position for each box at each transition), you get the
following results:
transition | CVF | approximate filter wheel position |
wavelength |
---|---|---|---|
Paschen beta | 0 | 6955 | 1282 nm |
Brackett gamma | 0 | 4955 | 2166 nm |
Brackett gamma | 1 | 3605 | 2166 nm |
Brackett beta | 1 | 2385 | 2626 nm |
Brackett beta | 2 | 1330 | 2626 nm |
parameter | oct98mlof | apr99wiro | oct99wiro | apr00wiro |
---|---|---|---|---|
0.slope | -439 | -440 | -445 | -440 |
0.L0 | 1176 | 1167 | 1160 | 1159 |
0.Lmax | 2282 | 2267 | 2274 | 2265 |
1.slope | -379 | -377 | -376 | -372 |
1.L0 | 1848 | 1849 | 1854 | 1860 |
1.Lmax | 2794 | 2790 | 2793 | 2791 |
2.slope | -900 | -900 | -900 | -900 |
2.L0 | 2318 | 2322 | 2326 | 2322 |
2.Lmax | 4573 | 4572 | 4578 | 4573 |
These numbers should be entered into the the filter wheel file. As of this writing (11 August 2000), we are using Filter Wheel V (V as in 5). The name of the file is filterV.fts, and it's probably in /applic/runtime/fwheels/ on the data computer. The filter wheel positions for the short-wavelength ends of the CVFs 0.Fw0, 1.Fw0, and 2.Fw0 (the numbers given in the last column of Table FW1) are also in this file --- never, ever, ever change these parameters in the file under any circumstances.
Finally, as mentioned above, we sometimes also spot-check the alignment of the filter wheel while on an observing run. One way to do this is to pick a broadband filter (the K-band filter is good for this) and walk through the filter at 10-step increments, computing a box mean or box total at each step. Plotting intensity v. filter wheel position should give a fairly symmetric, centrally-peaked function, the centroid of which is the central filter wheel position of that filter. The overall angular orientation of the filter wheel can then easily be updated in the software so that the center of the broadband corresponds to one of the following numbers (expected by the software): J at 8890, H at 8610, K at 8335, 3.3 at 8055, L'' at 7780, and M' at 7500.